Biomechanical analysis of a novel femoral neck locking plate for treatment of vertical shear Pauwel's type C femoral neck fractures

Peter J. Nowotarski, Bain Ervin, Brian Weatherby, Jonathan Pettit, Ron Goulet, Brent Norris

Research output: Contribution to journalArticle

38 Citations (Scopus)

Abstract

Background: The purpose of this study is to determine the biomechanical stability of a novel prototype femoral neck locking plate (FNLP) for treatment of Pauwels type C femoral neck fractures compared with other current fixation methods. Methods: Forty femur sawbones were divided into groups and a vertical femoral neck fracture was made. Each group was repaired with one of the following: (CS) three parallel cancellous screws; (XCS) two cancellous lag screws into the head and one transverse lag screw into the calcar; and (FNLP) a novel FNLP with two 5.7 mm locking, one lag screw into the calcar and two screws into the shaft; and (AMBI) a two-hole, 135° AMBI plate with a derotation screw. All groups were tested for change in axial stiffness over 20 000 cycles, and rotational stiffness was measured before and after cyclic testing. A maximum load to failure test was also conducted. Results were compared with one-way analysis of variance (ANOVA) and Fisher protected least significant difference (PLSD). Results: Results for axial stiffness show that AMBI, CS, XCS and FNLP are 2779.0, 2207.2, 3029.9 and 3210.7 N-m mm -1, respectively. Rotational rigidity results are 4.5, 4.1, 17.1 and 18.7 N-m mm -1. The average cyclic displacements were 0.75, 0.88, 0.80 and 0.65 mm, respectively. Destructive failure loads for AMBI, CS, XCS and FNLP were 2.3, 1.7, 1.6 and 1.9 kN, respectively. Conclusions: The results of this experiment show statistically significant increases in axial stiffness for the FNLP compared with three traditional fixation methods. The FNLP demonstrates increased mechanical stiffness and combines the desirable features of current fixation methods.

Original languageEnglish
Pages (from-to)802-806
Number of pages5
JournalInjury
Volume43
Issue number6
DOIs
StatePublished - 1 Jun 2012
Externally publishedYes

Fingerprint

Femoral Neck Fractures
Femur Neck
Femur
Analysis of Variance
Head

Keywords

  • Biomechanical
  • Cyclic
  • Hip
  • Locking
  • Plate

Cite this

Nowotarski, Peter J. ; Ervin, Bain ; Weatherby, Brian ; Pettit, Jonathan ; Goulet, Ron ; Norris, Brent. / Biomechanical analysis of a novel femoral neck locking plate for treatment of vertical shear Pauwel's type C femoral neck fractures. In: Injury. 2012 ; Vol. 43, No. 6. pp. 802-806.
@article{11536ed555414a2787e38079d47fa64d,
title = "Biomechanical analysis of a novel femoral neck locking plate for treatment of vertical shear Pauwel's type C femoral neck fractures",
abstract = "Background: The purpose of this study is to determine the biomechanical stability of a novel prototype femoral neck locking plate (FNLP) for treatment of Pauwels type C femoral neck fractures compared with other current fixation methods. Methods: Forty femur sawbones were divided into groups and a vertical femoral neck fracture was made. Each group was repaired with one of the following: (CS) three parallel cancellous screws; (XCS) two cancellous lag screws into the head and one transverse lag screw into the calcar; and (FNLP) a novel FNLP with two 5.7 mm locking, one lag screw into the calcar and two screws into the shaft; and (AMBI) a two-hole, 135° AMBI plate with a derotation screw. All groups were tested for change in axial stiffness over 20 000 cycles, and rotational stiffness was measured before and after cyclic testing. A maximum load to failure test was also conducted. Results were compared with one-way analysis of variance (ANOVA) and Fisher protected least significant difference (PLSD). Results: Results for axial stiffness show that AMBI, CS, XCS and FNLP are 2779.0, 2207.2, 3029.9 and 3210.7 N-m mm -1, respectively. Rotational rigidity results are 4.5, 4.1, 17.1 and 18.7 N-m mm -1. The average cyclic displacements were 0.75, 0.88, 0.80 and 0.65 mm, respectively. Destructive failure loads for AMBI, CS, XCS and FNLP were 2.3, 1.7, 1.6 and 1.9 kN, respectively. Conclusions: The results of this experiment show statistically significant increases in axial stiffness for the FNLP compared with three traditional fixation methods. The FNLP demonstrates increased mechanical stiffness and combines the desirable features of current fixation methods.",
keywords = "Biomechanical, Cyclic, Hip, Locking, Plate",
author = "Nowotarski, {Peter J.} and Bain Ervin and Brian Weatherby and Jonathan Pettit and Ron Goulet and Brent Norris",
year = "2012",
month = "6",
day = "1",
doi = "10.1016/j.injury.2011.09.012",
language = "English",
volume = "43",
pages = "802--806",
journal = "Injury",
issn = "0020-1383",
publisher = "Elsevier Ltd",
number = "6",

}

Biomechanical analysis of a novel femoral neck locking plate for treatment of vertical shear Pauwel's type C femoral neck fractures. / Nowotarski, Peter J.; Ervin, Bain; Weatherby, Brian; Pettit, Jonathan; Goulet, Ron; Norris, Brent.

In: Injury, Vol. 43, No. 6, 01.06.2012, p. 802-806.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Biomechanical analysis of a novel femoral neck locking plate for treatment of vertical shear Pauwel's type C femoral neck fractures

AU - Nowotarski, Peter J.

AU - Ervin, Bain

AU - Weatherby, Brian

AU - Pettit, Jonathan

AU - Goulet, Ron

AU - Norris, Brent

PY - 2012/6/1

Y1 - 2012/6/1

N2 - Background: The purpose of this study is to determine the biomechanical stability of a novel prototype femoral neck locking plate (FNLP) for treatment of Pauwels type C femoral neck fractures compared with other current fixation methods. Methods: Forty femur sawbones were divided into groups and a vertical femoral neck fracture was made. Each group was repaired with one of the following: (CS) three parallel cancellous screws; (XCS) two cancellous lag screws into the head and one transverse lag screw into the calcar; and (FNLP) a novel FNLP with two 5.7 mm locking, one lag screw into the calcar and two screws into the shaft; and (AMBI) a two-hole, 135° AMBI plate with a derotation screw. All groups were tested for change in axial stiffness over 20 000 cycles, and rotational stiffness was measured before and after cyclic testing. A maximum load to failure test was also conducted. Results were compared with one-way analysis of variance (ANOVA) and Fisher protected least significant difference (PLSD). Results: Results for axial stiffness show that AMBI, CS, XCS and FNLP are 2779.0, 2207.2, 3029.9 and 3210.7 N-m mm -1, respectively. Rotational rigidity results are 4.5, 4.1, 17.1 and 18.7 N-m mm -1. The average cyclic displacements were 0.75, 0.88, 0.80 and 0.65 mm, respectively. Destructive failure loads for AMBI, CS, XCS and FNLP were 2.3, 1.7, 1.6 and 1.9 kN, respectively. Conclusions: The results of this experiment show statistically significant increases in axial stiffness for the FNLP compared with three traditional fixation methods. The FNLP demonstrates increased mechanical stiffness and combines the desirable features of current fixation methods.

AB - Background: The purpose of this study is to determine the biomechanical stability of a novel prototype femoral neck locking plate (FNLP) for treatment of Pauwels type C femoral neck fractures compared with other current fixation methods. Methods: Forty femur sawbones were divided into groups and a vertical femoral neck fracture was made. Each group was repaired with one of the following: (CS) three parallel cancellous screws; (XCS) two cancellous lag screws into the head and one transverse lag screw into the calcar; and (FNLP) a novel FNLP with two 5.7 mm locking, one lag screw into the calcar and two screws into the shaft; and (AMBI) a two-hole, 135° AMBI plate with a derotation screw. All groups were tested for change in axial stiffness over 20 000 cycles, and rotational stiffness was measured before and after cyclic testing. A maximum load to failure test was also conducted. Results were compared with one-way analysis of variance (ANOVA) and Fisher protected least significant difference (PLSD). Results: Results for axial stiffness show that AMBI, CS, XCS and FNLP are 2779.0, 2207.2, 3029.9 and 3210.7 N-m mm -1, respectively. Rotational rigidity results are 4.5, 4.1, 17.1 and 18.7 N-m mm -1. The average cyclic displacements were 0.75, 0.88, 0.80 and 0.65 mm, respectively. Destructive failure loads for AMBI, CS, XCS and FNLP were 2.3, 1.7, 1.6 and 1.9 kN, respectively. Conclusions: The results of this experiment show statistically significant increases in axial stiffness for the FNLP compared with three traditional fixation methods. The FNLP demonstrates increased mechanical stiffness and combines the desirable features of current fixation methods.

KW - Biomechanical

KW - Cyclic

KW - Hip

KW - Locking

KW - Plate

UR - http://www.scopus.com/inward/record.url?scp=84860753706&partnerID=8YFLogxK

U2 - 10.1016/j.injury.2011.09.012

DO - 10.1016/j.injury.2011.09.012

M3 - Article

C2 - 22019259

AN - SCOPUS:84860753706

VL - 43

SP - 802

EP - 806

JO - Injury

JF - Injury

SN - 0020-1383

IS - 6

ER -